Patients may require all or part of
their respiratory support to be provided by a mechanical ventilator. Weaning from
mechanical ventilation may follow several patterns. In patients
ventilated for short periods (no more than a few days), it is common to allow
20–30min breathing on a ‘T’ piece before removing the endotracheal tube.
For patients who have received longer term ventilation, it is unlikely
that mechanical support can be
withdrawn suddenly; several methods are
commonly used to wean these patients from mechanical ventilation.
There is no strong evidence that any technique is superior in terms of
weaning success.
Intermittent
‘T’ piece or continuous positive airway pressure
(CPAP)
Spontaneous breathing is allowed for
increasingly prolonged periods with a rest on mechanical ventilation in
between. The use of a ‘T’ piece for longer than 30min may lead to basal
atelectasis since the endotracheal tube bypasses the physiological PEEP
effect of the larynx. Therefore, it is common to use 5cmH2O CPAP as spontaneous breathing periods
get longer. In the early stages of
weaning, mechanical ventilation is often continued at night to encourage sleep,
avoid fatigue, and rest respiratory muscles.
Intermittent
mandatory ventilation (IMV)
The set mandatory rate is gradually
reduced as the spontaneous rate increases. Spontaneous breaths are
usually pressure-supported to overcome circuit and ventilator valve resistance.
With this technique, it is important that the patient’s required
minute ventilation is provided by the combination of mandatory breaths and
spontaneous breaths without an excessive spontaneous rate. The
reduction in mandatory rate should be
slow enough to maintain adequate minute
ventilation. It is also important that the patient can synchronise his own
respiratory efforts with mandatory ventilator breaths; many cannot, particularly
where there are frequent spontaneous breaths, some of
which may ‘stack’ with mandatory breaths causing hyperinflation.
Pressure
support ventilation
All respiratory efforts are spontaneous
but positive pressure is added to each breath, the level being chosen
to maintain an appropriate tidal volume. Weaning is performed by a
gradual reduction of the pressure support level while the respiratory rate
is <30/min. The patient is extubated or allowed to breathe with 5cmH2O CPAP when pressure support
is minimal (<10–15cmH2O with modern ventilators).
Choice
of ventilator
Modern ventilators have enhancements to
aid weaning; however, weaning most patients from ventilation is
possible with a basic ventilator and the intermittent ‘T’ piece technique,
provided an adequate fresh gas flow is provided. If IMV and/or pressure support
are used, the ventilator should provide the features listed in the table opposite.
Key
features in the choice of ventilator
• Ventilator must allow patient triggering
(i.e. not a minute volume divider).
• Fresh gas flow must be greater than
spontaneous peak inspiratory flow.
• Minimum circuit resistance (short, wide
bore, and smooth internal lumen).
• Low resistance ventilator valves.
• Sensitive pressure or flow trigger (ideally
monitored close to the ET tube).
• Synchronised IMV (avoids ‘stacking’
mandatory on spontaneous breaths).
0 comments:
Post a Comment