Failure
to ventilate despite apparently appropriate ventilator settings and exclusion
of patient factors requires assessment of which ventilator settings are not
being delivered.
Poor gas exchange
• Increase
FIO2 to 1.0 and start manual ventilation.
• Exclude
patient factors.
• Check
endotracheal tube is correctly positioned, patent, and both lungs are being
inflated. Consider tube replacement.
• Check
ventilator circuit is both intact and patent, and ventilator is functioning
correctly. Check ventilator settings, including FIO2,
PEEP, I:E ratio, set VT,
respiratory rate and/or pressure control.
Low VT or low pressure alarm
Expired
VT will be lower than inspired VT
with bronchopleural fistula. In other situations where
expired VT is suddenly
significantly lower than inspired VT, or
maintenance of circuit pressure is not possible, there is a leak in the
ventilator circuit. The patient should be manually ventilated while the leak is
identified and corrected. If the leak persists with manual ventilation,
repositioning or replacement of the endotracheal tube is required.
High pressure alarm
A
sudden increase in airway pressure indicates a change in resistance to gas flow.
After patient factors have been excluded:
• Check
patency of the endotracheal tube (e.g. with a suction catheter); re-intubate if
in doubt.
• Check
patency of ventilator circuit and catheter mount. Look for excessive trapped
water partially occluding the catheter mount.
• Check
for sputum obstructing the HME.
• Consider
a chest X-ray to identify endotracheal tube malposition (e.g. cuff above vocal
cords, tip at carina, tube in main bronchus).
0 comments:
Post a Comment