2/7/14

Blood pressure measurement

Monitoring a client’s blood pressure forms part of the assessment of vital signs and assists nursing and medical staff in determining the client’s cardiovascular status, general health and well-being. The only pieces of equipment needed are an appropriate device and a stethoscope (if required).

It is recommended that you use an electronic device for infants and children as the pulse is generally difficult to hear. The most commonly used devices are:

• manual mercury sphygmomanometer, which measures blood pressure using mercury
• manual aneroid sphygmomanometer, which measures blood pressure using air
• electronic automated devices, which measure the sound of the blood as it passes through the artery, and then display the pressure on a monitor.

All the devices measure in millimetres of mercury, usually expressed as mmHg. Remember, machines can and do malfunction; if the reading seems too high or low given the client’s overall condition and demeanour recheck the measurement either by using another machine or by checking it manually. All devices should be inspected and calibrated annually.

The procedures are outlined below.
Procedure
Rationale
Explain procedure and ensure adequate
understanding
To ensure informed consent and obtain
client’s co-operation
Before proceeding check that the client
is rested, and has not been consuming
alcohol or nicotine, and advise them not
to talk during the procedure
To ensure an accurate reading
Ensure client is positioned correctly
To ensure accuracy of reading. Blood pressure
readings should be taken in the same
position and on the same limb each time if
possible. The dominant side is recommended
if at all possible. It is therefore good practice
to document position to ensure continuity of
care, e.g. lying, standing, sitting, and whether
you have used the left or right limb. In some
instances it may be necessary to record both
lying and standing blood pressure, in which
case it is important to distinguish between
the two as per local preferences
Ensure correct-size cuff
Correct cuff size must be used for accurate
reading. Size and instructions are usually
printed on the inside of cuff
Apply the cuff 2.5 cm above the antecubital
fossa with the client’s palm upwards,
ensuring that the cuff is level with the
sphygmomanometer and the client’s heart
To ensure accuracy of reading. Inaccurate
readings occur if equipment is not in correct
alignment. NB Do not use a limb receiving
intravenous therapy and make sure that
clothing above the cuff is not restrictive
otherwise the accuracy of the reading will
be questionable
If using an automated device commence
the reading
This reading provides a rough estimate of
the systolic blood pressure and reduces the
possible discomfort arising from hyper
inflation of the cuff
If undertaking the measurement manually,
locate the radial pulse on the cuffed arm
then inflate the cuff until the pulse is no
longer felt – note the point at which this
happens on the sphygmomanometer scale

Deflate the cuff for 30 seconds ensuring
that all air has been released
To allow circulation to return to the limb
Locate the brachial artery and place the
stethoscope over the pulse
Locating the artery prior to inflating the
cuff ensures that the strongest point of the
pulse can be identified and heard during the
reading
Inflate the cuff again to approximately
20–30mmHg above the estimated systolic
reading
To apply pressure to the artery in order to
then hear and record the blood pressure
Deflate the cuff, slowly, at approximately
2–3mmHg per second and listen
Whilst slow deflation may be uncomfortable
for the client, deflating too quickly will not
allow time to hear the blood pressure
sounds and mean that the procedure has to
be repeated
Listen for and note the first sound that
you hear (systolic); this will be followed
by the other sounds listed. Note the
point at which the sound disappears
(diastolic)
To the attuned, five different sounds can be
heard, known as Korotkoff sounds:
1 The first clear tapping sound = systolic
2 A swishing sound
3 An intense clear tap
4 A muffled sound
5 Sounds disappear altogether = diastolic
Compare reading with client’s previous
reading and normal range and note any
abnormality or improvement. If necessary
repeat the reading but allow the
client to rest the limb between attempts
Normal range:
Newborn                   80/40 mmHg
1–7 years                  100/65 mmHg
8–12 years                100/70 mmHg
Adult                          120/80 mmHg
Elderly                       systolic 100 + age
                                  In the elderly the diastolic
                                  may rise also
Record blood pressure measurement on
appropriate documentation (see Figure
4.1) and report any abnormalities
Legal requirement to maintain documentation
and safeguard client through effective
communications
Remove and clean equipment
To reduce the risk of cross-infection.
NB Automated machines should not be
used on clients being nursed in isolation
Wash hands thoroughly

0 comments:

Post a Comment