It is recommended that you use an electronic device for infants and children as the pulse is generally difficult to hear. The most commonly used devices are:
• manual mercury sphygmomanometer, which measures blood pressure using mercury
• manual aneroid sphygmomanometer, which measures blood pressure using air
• electronic automated devices, which measure the sound of the blood as it passes through the artery, and then display the pressure on a monitor.
All the devices measure in millimetres of mercury, usually expressed as mmHg. Remember, machines can and do malfunction; if the reading seems too high or low given the client’s overall condition and demeanour recheck the measurement either by using another machine or by checking it manually. All devices should be inspected and calibrated annually.
The procedures are outlined below.
Procedure
|
Rationale
|
Explain procedure
and ensure adequate
understanding
|
To ensure informed consent and obtain
client’s co-operation
|
Before proceeding
check that the client
is rested, and has
not been consuming
alcohol or nicotine,
and advise them not
to talk during the
procedure
|
To ensure an accurate reading
|
Ensure client is
positioned correctly
|
To ensure accuracy of reading. Blood pressure
readings should be taken in the same
position and on the same limb each time if
possible. The dominant side is recommended
if at all possible. It is therefore good practice
to document position to ensure continuity of
care, e.g. lying, standing, sitting, and whether
you have used the left or right limb. In some
instances it may be necessary to record both
lying and standing blood pressure, in which
case it is important to distinguish between
the two as per local preferences
|
Ensure correct-size
cuff
|
Correct cuff size must be used for accurate
reading. Size and instructions are usually
printed on the inside of cuff
|
Apply the cuff 2.5
cm above the antecubital
fossa with the
client’s palm upwards,
ensuring that the
cuff is level with the
sphygmomanometer and
the client’s heart
|
To ensure accuracy of reading. Inaccurate
readings occur if equipment is not in correct
alignment. NB Do
not use a limb receiving
intravenous therapy and make sure that
clothing above the cuff is not restrictive
otherwise the accuracy of the reading will
be questionable
|
If using an
automated device commence
the reading
|
This reading provides a rough estimate of
the systolic blood pressure and reduces the
possible discomfort arising from hyper
inflation of the cuff
|
If undertaking the
measurement manually,
locate the radial
pulse on the cuffed arm
then inflate the
cuff until the pulse is no
longer felt – note
the point at which this
happens on the
sphygmomanometer scale
|
|
Deflate the cuff for
30 seconds ensuring
that all air has
been released
|
To allow circulation to return to the limb
|
Locate the brachial
artery and place the
stethoscope over the
pulse
|
Locating the artery prior to inflating the
cuff ensures that the strongest point of the
pulse can be identified and heard during the
reading
|
Inflate the cuff
again to approximately
20–30mmHg above the
estimated systolic
reading
|
To apply pressure to the artery in order to
then hear and record the blood pressure
|
Deflate the cuff,
slowly, at approximately
2–3mmHg per second
and listen
|
Whilst slow deflation may be uncomfortable
for the client, deflating too quickly will not
allow time to hear the blood pressure
sounds and mean that the procedure has to
be repeated
|
Listen for and note
the first sound that
you hear (systolic);
this will be followed
by the other sounds
listed. Note the
point at which the
sound disappears
(diastolic)
|
To the attuned, five different sounds can be
heard, known as Korotkoff sounds:
1 The first clear tapping sound = systolic
2 A swishing sound
3 An intense clear tap
4 A muffled sound
5 Sounds disappear altogether = diastolic
|
Compare reading with
client’s previous
reading and normal
range and note any
abnormality or
improvement. If necessary
repeat the reading
but allow the
client to rest the
limb between attempts
|
Normal range:
Newborn
80/40 mmHg
1–7 years
100/65 mmHg
8–12 years
100/70 mmHg
Adult 120/80 mmHg
Elderly systolic 100 + age
In the
elderly the diastolic
may rise
also
|
Record blood
pressure measurement on
appropriate
documentation (see Figure
4.1) and report any
abnormalities
|
Legal requirement to maintain documentation
and safeguard client through effective
communications
|
Remove and clean
equipment
|
To reduce the risk of cross-infection.
NB Automated machines should not be
used on clients being nursed in isolation
|
Wash hands thoroughly
|
|
0 comments:
Post a Comment