1/1/14

Troubleshooting Ventilator Alarms

Troubleshooting Ventilator Alarms

■ When the ventilator alarms: Check the Pt first. If Pt is in no apparent distress, check ventilator to determine source of problem.
■ If patient is showing signs of distress (“fighting the vent”): Try to calm the Pt. If unsuccessful, immediately disconnect Pt from vent and manually ventilate with 100% oxygen using a BVM. Notify the physician and RT immediately.

Alarm
Intervention
Low-Pressure:
Usually caused
by system
disconnections
or leaks.
Reconnect Pt to ventilator.
Evaluate cuff and reinflate if needed (if ruptured, tube will need to be replaced).
Evaluate connections and tighten or replace as needed.
Check ET tube placement (auscultate lung fields and assess for equal, bilateral breath sounds).
High-Pressure:
Usually caused by
resistance within
the system. Can
be kink or water
in tubing, Pt biting
the tube, copious
secretions, or
plugged endotracheal
tube.
Suction Pt if secretions are suspected.
Insert bite block to prevent Pt from biting tube.
Reposition Pt’s head and neck, or reposition tube.
Sedation may be required to prevent a Pt from fighting the vent, but only after
careful assessment excludes a physical or mechanical cause.
High Respiratory
Rate:
Can be caused by
anxiety or pain,
secretions in
ETT/airway, or
hypoxia.
Suction Pt.
Look for source of anxiety (e.g., pain, environmental stimuli, inability to communicate, restlessness, etc.).
Evaluate oxygenation.
Low Exhaled
Volume:
Usually caused
by tubing disconnection
or
inadequate seal.
Evaluate/reinflate cuff; if ruptured, ETT must be replaced.
Evaluate connections; tighten or replace as needed; check ETT placement,
reconnect to ventilator.


0 comments:

Post a Comment